Table 3 Descriptive data (mean ± SD), median, and quartile range for the tested light-curing tips comparing old and new light-curing unit tips (Wilcoxon Two-Sample Test)

From: Effect of damage or contamination to the tips of 200 light-curing units

Variable

 

Old tip (n = 154)

New tips (n = 46)

p-value

Contamination Score

(n = 62)

Mean

Standard Deviation

Median

Quartile Range

0.87

1.85

0

1.00

1.30

2.19

0

2

0.091

Physical Damage Score

(n = 35)

Mean

Standard Deviation

Median

Quartile Range

0.64

1.66

0

0

0.85

2.20

0

0

0.93

Power Reduction (%)

Mean

Standard Deviation

Median

Quartile Range

20.5

18.07

15.0

26.34

15.1

15.57

13.35

22.85

0.059

Radiant Emittance Reduction (%)

Mean

Standard Deviation

Median

Quartile Range

10.0

17.32

0

13.52

7.9

13.39

0

10.28

0.87